Internal Quality Control Kits for Immunochromatographic Test SD Bioline Assay/Capillia/TBcID

A positive and a negative control must be tested with each new lot or a new shipment of kits received and with each new batch of extraction buffer prepared. Similarly, these controls must be run weekly or along with each batch of patient isolates when tests are set up less frequently.

1. Frequency:

Reading MGIT 960 Tubes Manually

The beauty of the MGIT 960 tubes is that they can be read by the instrument and also manually. There are times AFB cultures are incubated at temperatures other than 37oC. If that is the case, the MGIT tube can be incubated at an alternate temperature. These tubes would then need to be read manually with a Wood’s Lamp or other UV sources of the proper wavelength and intensity.