Content Status

Type

Linked Node

  • Maintenance of CBNAAT Instrument under NTEP

    Learning Objectives

    Maintenance of CBNAAT Instrument under NTEP

Content

Maintenance of the CBNAAT instrument is an essential activity to be performed in the laboratory setting and involves:

Preventive Maintenance Tasks

  • Performed regularly by the user
  • Ensure the good performance of the system
  • Avoid problems of malfunction

Need-based Maintenance Tasks

  • Performed in specific situations
  • Performed or guided by manufacturer representatives

 

Frequency-based Tasks for the CBNAAT Instrument

DAILY WEEKLY MONTHLY ANNUAL OR AFTER 2000 TESTS PER MODULE
  • Remove and properly dispose off cartridges
  • Clean and disinfect the work area
  • Put on a dust cover when the instrument is not in use
  • Disinfect the instrument’s surface
  • Disinfect cartridge bay interior and plunger rod
  • Filter cleaning

 

  • Disinfect cartridge bay interior and plunger rod
  • Clean instrument filter 
  • Archive and back-up test results
  • Filter cleaning

 

Module Xpert Check (calibration) and maintenance of the module  

 

Materials Required for CBNAAT Instrument  Maintenance

  • Freshly prepared 1% sodium hypochlorite solution
  • 70% Ethanol
  • Wipes, tissues, or cotton
  • Disposable gloves 
  • Clean water and soap (for washing the filters)
  • Replacement filters for the fan (available from the manufacturer). 

 